| General Specifications |
| Temperature |
| |
-40 to 550 °C (-40 to 1022 ° F) |
|
| Display Resolution |
| |
0.1°C (0.1 °F) of reading |
|
| D:S (Distance to spot size) |
|
| Easy Emissivity Selector |
| |
Adjustable with three settings: |
| |
Low (0.3), Medium (0.7). High (0.95) |
|
| Display Accuracy [Assumes ambient operating temperature of 23°C(73°F) to 25°C(77°F)] |
| |
±1.0 % of reading ±1 % of reading or ±1 °C (±2 °F), whichever is greater below 0 °C/32 °F, ±1 °C (±2 °F) ± 0.1°/1° |
|
| Response Time |
| |
500 mSec (95 % of reading) |
|
| Repeatability |
| |
±0.5 % of reading or ±1 °C (±2 °F), whichever is greater |
|
| Spectral Response |
|
| Laser Sighting |
|
| Laser Shutoff |
| |
Laser turns off above ambient temperature of 40 °C (104 °F) |
|
| Laser Power |
| |
Class 2(II) operation; Output <1mW, wavelength 630-670 nm |
|
| Relative Humidity |
| |
10 % to 90 % RH non-condensing, at <30 °C (86 °F) |
|
| Power, battery life |
| |
2 AA batteries (alkaline or NiCD) |
|
| Battery Life |
|
| Display Hold |
|
| Backlit Display |
| |
Yes, LCD with dual temperatures (current and MAX/MIN/DIF/KTC), low battery, F/C indicator, and Scan/Hold options |
|
| Warranty |
|
| Operating Temperature |
| |
0 °C to 50 °C (32 °F to 120 °F) |
|
| Storage Temperature |
| |
-20 °C to 65 °C (-25 °F to 150 °F) |
|
| MAX, MIN, DIF Temperatures |
|
| Thermocouple Type K Mini Adapter Input |
| |
Yes, compatible with industry standard K type probes with mini-connector. Displays to the thermometer temperature range. |
|
| Thermocouple Type K Fabric Hook-and-Loop Fastener Pipe Probe |
| |
Yes, with a temperature range of 0 °C to 100 °C (32 °F to 212 °F) and accuracy of ±2.2 °C (±4 °F) |
|
| HVAC Measurement Guide |
|